Related Experiment Video
Updated: Jul 30, 2026

A Rat Model of Ventricular Fibrillation and Resuscitation by Conventional Closed-chest Technique
Published on: April 26, 2015
The effects of biphasic and conventional monophasic defibrillation on postresuscitation myocardial function
1The Institute of Critical Care Medicine, Palm Springs, California 92262-5309, USA.
Objectives:
The purpose of this study was to compare the effects of biphasic defibrillation waveforms and conventional monophasic defibrillation waveforms on the success of initial defibrillation, postresuscitation myocardial function and duration of survival after prolonged ventricular fibrillation (VF).
Background:
We have recently demonstrated that the severity of postresuscitation myocardial dysfunction was closely related to the magnitude of the electrical energy of the delivered defibrillation shock. In the present study, the effects of fixed 150-J low-energy biphasic waveform shocks were compared with conventional monophasic waveform shocks after prolonged VF.
Methods:
Twenty anesthetized, mechanically ventilated domestic pigs were investigated. VF was induced with an AC current delivered to the right ventricular endocardium. After either 4 or 7 min of untreated ventricular fibrillation (VF), the animals were randomized for attempted defibrillation with up to three 150-J biphasic waveform shocks or conventional sequence of 200-, 300- or 360-J monophasic waveform shocks. If VF was not reversed, a 1-min interval of precordial compression preceded a second sequence of up to three shocks. The protocol was repeated until spontaneous circulation was restored or for a total of 15 min.
Results:
Monophasic waveform defibrillation after 4 or 7 min of untreated VF resuscitated eight of 10 pigs. All 10 pigs treated with biphasic waveform defibrillation were successfully resuscitated. Transesophageal echo-Doppler, arterial pressure and heart rate measurements demonstrated significantly less impairment of cardiovascular function after biphasic defibrillation.
Conclusions:
Lower-energy biphasic waveform shocks were as effective as conventional higher energy monophasic waveform shocks for restoration of spontaneous circulation after 4 and 7 min of untreated VF. Significantly better postresuscitation myocardial function was observed after biphasic waveform defibrillation.
Related Concept Videos
Electrophysiology of Normal Cardiac Rhythm
Cardiac Action Potential
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials
Dysrhythmias VI: Management of Dysrhythmias
Cardiopulmonary Resuscitation III: AED Use
Cardiopulmonary Resuscitation IV: Pharmacological Management
Cardiomyopathy V: Interprofessional Care