Objectives:
To determine the dose response function for the Digora(R) (Soredex Orion Corporation, Helsinki, Finland) storage phosphor system for intra-oral radiography.
Methods:
Specially designed software provided by the manufacturer was used to register data from the pre-scanning that takes place prior to readout. These data contain information about the signal from the photomultiplier and the amplification that forms the basis for the AD conversion to 8 bit image data. Dose response functions were determined for different calibration settings and also combined into one single function.
Results:
The dose response function is linear irrespective of calibration. The gray levels of the final radiographs are directly proportional to the amplified signal. The Digora(R) does not fully compensate for exposure differences. Increased exposure results in slightly darker radiographs.
Conclusions:
Since the gray levels of the final radiographs are directly proportional to the amplified signal, the gray levels may be used as a relative measure of the exposure in the Digora(R) system provided that they are subtracted from the maximum value of 255.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:48High Spatial Resolution Chemical Imaging of Implant-Associated Infections with X-ray Excited Luminescence Chemical Imaging Through Tissue
Published on: September 30, 2022
Related Concept Videos
X-ray Imaging
Dose-Response Relationship: Overview
Photoluminescence: Applications
Determination of Crystal Structures
Pharmacokinetic–Pharmacodynamic Relationship: Duration of Dose-Effect Relationship
Dose Response Curve: Conventional Versus Nonmonotonic