Related Experiment Videos
Establishing a quality control program for an automated dosimetry system
1Department of Radiology, University of Minnesota, Minneapolis 55455, USA.
Abstract:
Automatic dosimetry systems can provide instantaneous dose and dose-rate information during fluoroscopic procedures as well as long-term records of patient doses. For this information to be useful, it is necessary that the accuracy of such systems be maintained through a rigorous quality control program. Daily and weekly quality control checks were performed on a PEMNET automated dosimetry system to determine its stability and the value of such tests in a quality control program. Weekly tests included monitoring the accuracy of the measured doses under a variety of conditions. The results of the tests indicate possible improvements in test methodology and real and potential sources of system failure and provide a statistical basis for setting quality control limits for future system monitoring.