Related Experiment Videos
Real-time imaging and triggering of 3D contrast-enhanced MR angiograms using MR fluoroscopy
S J Riederer1, S B Fain, D G Kruger
1Magnetic Resonance Laboratory, Mayo Clinic, Rochester, MN 55905, USA. riederer@mayo.edu
Abstract:
Although a variety of timing techniques and data acquisition strategies have been used for three-dimensional contrast-enhanced MR angiography, many are still limited by inadequate overall reliability, limited spatial resolution, or complexity. A technique is presented in this work in which contrast arrival is detected in the targeted vasculature in real time using MR fluoroscopy. Upon detection the operator triggers a 3D MR angiographic acquisition which uses an elliptical centric view order. It is shown that the view order intrinsically provides a high degree of venous suppression which in turn allows acquisition times of 30 s or longer, permitting high spatial resolution. The reliability of fluoroscopic triggering in bolus detection is shown to be over 90%. The technique provides high quality contrast-enhanced MR angiograms for many vascular regions.