Related Experiment Videos
Long-term clinical experience with the Omnicarbon prosthetic valve
S Torregrosa1, J Gómez-Plana, F J Valera
1Servicio de Cirugía Cardiovascular, Hospital Universitario La Fé, Valencia, Spain.
The Annals of Thoracic Surgery
|October 6, 1999
Summary
The Omnicarbon valve prosthesis demonstrated excellent clinical outcomes over a 10-year follow-up. Survival rates were high, with minimal structural failure and low rates of complications like hemorrhage and thromboembolism.
Area of Science:
- Cardiovascular Surgery
- Biomaterials Science
- Clinical Outcomes Research
Background:
- A total of 781 Omnicarbon valve prostheses were implanted in 647 patients between February 1985 and December 1994.
- The patient cohort included 357 males and 290 females with a mean age of 53.5 years.
- Prior to surgery, 81% of patients were in New York Heart Association (NYHA) class III or IV, indicating severe cardiac dysfunction.
Purpose of the Study:
- To evaluate the long-term clinical performance and safety of the Omnicarbon valve prosthesis.
- To assess survival rates, complication incidence, and functional status following valve replacement surgery.
Main Methods:
- The study involved 227 aortic valve replacements (AVR), 286 mitral valve replacements (MVR), and 134 double-valve replacements (DVR).
- Follow-up data was 96.3% complete, accumulating 2,746 patient-years with a maximum follow-up of 10.7 years.
Main Results:
- Hospital mortality rates ranged from 7.0% (AVR) to 8.2% (DVR).
- Annualized complication rates included 0.8% for hemorrhage and 0.7% for thromboembolism.
- No structural valve failure was observed; nonstructural dysfunction occurred at 0.8% per patient-year, with dehiscence being the most common cause.
- Prosthetic valve endocarditis occurred at 0.3% per patient-year.
- At 10 years, 91% of survivors were in NYHA class I or II.
- Overall 10-year survival was 82.5%; valve-related mortality yielded a 10-year survival of 91.9%.
Conclusions:
- The Omnicarbon prosthesis exhibits excellent clinical results over a decade of follow-up.
- High survival rates and good functional recovery are achieved.
- The prosthesis demonstrates a favorable safety profile with low rates of major complications.