The role of image-guidance systems for head and neck surgery
R Metson1, M Cosenza, R E Gliklich
1Department of Otolaryngology, Massachusetts Eye and Ear Infirmary, Boston, USA.
Background:
Although image-guidance systems have gained widespread acceptance for neurosurgical procedures, their role for extracranial surgery of the head and neck is yet to be defined.
Objectives:
To describe the authors' experience with image-guidance systems and to measure the effects of image-guided technology on the performance of minimally invasive otolaryngological procedures.
Design:
Prospective cohort study.
Methods:
Optical- and electromagnetic-based image-guidance systems were used during the performance of endoscopic surgery on patients with disease of the paranasal sinuses, orbit, skull base, and temporal bone (n = 79). Results were compared with those in control patients who underwent similar surgery without image guidance during the same period (n = 42).
Results:
Intraoperative anatomical localization was accurate to within 2 mm at the start of surgery in all cases. Accuracy degraded by 0.89 +/- 0.20 mm (mean +/- SE) during the operative procedure. The use of an image-guidance system increased operating room time by a mean of 17.4 minutes per case (image-guidance group, 137.3 +/- 6.0 minutes [mean +/- SE]; control group, 119.9 +/- 5.7 minutes; P=.006) and increased hospital charges by approximately $496 per case. Intraoperative blood loss (image-guidance group, 178.4 +/- 18.0 mL [mean +/- SE]; control group, 149.4 +/- 20.1 mL) and complication rates (image-guidance group, 2.7%; control group, 4.7%) did not differ significantly between groups.
Conclusions:
Image-guidance systems can provide the head and neck surgeon with accurate information regarding anatomical localization in cases with poor surgical landmarks caused by extensive disease or prior surgery; however, the use of such systems is associated with increased operative time and expense.