Can reduced imaging times be used for scintimammography?
E A Clarke1, A Notghi, L K Harding
1Department of Physics and Nuclear Medicine, City Hospital NHS Trust, Birmingham, UK.
Abstract:
A reduction in acquisition times from 10 to 5 min for scintimammography does not reduce the diagnostic value of the study when imaging for detection of breast lesions. The test showed an overall sensitivity of 96%, specificity of 100% and accuracy of 98% for breast lesions visualized on both 5 and 10 min acquisitions. Even if lymph node detection is the primary concern of the study, longer scan times do not increase the sensitivity of the test (40% on both 5 and 10 min). Scintimammography is poor at detecting lymph nodes (sensitivity 40%, specificity 69-82%) and is not useful for assessing lymph node involvement. A reduction in imaging times appeared to be consistent on both camera systems tested, which have very different display outputs. This would indicate that other departments may be able to reduce times on their systems without affecting the quality of the study. The detection of breast lesions was also consistent between reporters.
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...