Related Experiment Videos
Evaluation of the automated coagulation analyzer SYSMEX CA 6000
P Quehenberger1, S Kapiotis, S Handler
1Clinical Institute of Medical and Chemical Laboratory Diagnostics, University of Vienna, Austria.
Thrombosis Research
|November 30, 1999
Summary
The SYSMEX CA 6000 analyzer offers precise results for routine and specialized coagulation tests, including single factor activities. Its cap-piercing system does not negatively impact performance, making it reliable for laboratory diagnostics.
Area of Science:
- Clinical diagnostics
- Hematology
- Laboratory automation
Background:
- Coagulation testing is crucial for diagnosing and monitoring hemostatic disorders.
- Evaluating new analytical platforms is essential for improving laboratory efficiency and accuracy.
- The SYSMEX CA 6000 is a modern coagulation analyzer with advanced features like photooptical clot detection and a cap-piercing system.
Purpose of the Study:
- To evaluate the technical performance of the SYSMEX CA 6000 coagulation analyzer.
- To assess its accuracy and precision for standard coagulation tests and single factor activities.
- To investigate potential interferences and carryover effects.
Main Methods:
- Evaluation of SYSMEX CA 6000 for prothrombin time, activated partial thromboplastin time, thrombin time, fibrinogen, and antithrombin.
- Determination of coagulation single factor activities (V, VII, VIII, IX).
- Interference studies with lipemic, icteric, and hemolytic samples; carryover studies with heparinized plasma.
Main Results:
- Most intra- and interassay coefficients of variation (CV) were below 5%, with minor exceptions noted for specific parameters.
- Minor interferences were observed with lipemic, icteric, and hemolytic samples, comparable to mechanical clot detection methods.
- No carryover was detected between heparinized and normal plasma samples.
- Excellent correlation (r=0.977-0.993) was found between SYSMEX CA 6000 and STA for factor activity measurements.
Conclusions:
- The SYSMEX CA 6000 analyzer demonstrates satisfactory precision for routine and specialized coagulation testing.
- The integrated cap-piercing system does not adversely affect the analyzer's performance.