ICP measurement control: laboratory test of 7 types of intracranial pressure transducers
M H Morgalla1, H Mettenleiter, M Bitzer
1Division of Neurosurgery, University of Tuebingen, Germany.
Abstract:
Intracranial pressure (ICP) monitoring has become an important parameter in the assessment of comatose patients, with raised intracranial pressure. The transducers in use have to fulfill the criteria of measurement accuracy, practicability and cost-effectiveness. However, these requirements are not always met in clinical practice. The need for ongoing quality control through independent laboratories remains. We have developed a laboratory set-up for the evaluation of intracranial pressure probes. Seven different types of currently used transducers have been tested for measurement accuracy. Under in vitro conditions 3 parameters were assessed: measurement accuracy, a 24 h drift and 10 day drifts. Tests for measurement accuracy were performed at increasing pressure levels of up to 80 mmHg. They were repeated 10 times per probe. This test allowed the simultaneous assessment of 5 different ICP probes. Drift was evaluated for 24 h and 10 days, at 6 pressure levels between 0 and 50 mmHg. Seven different types of ICP probes were tested (HanniSet, Camino, Codman, Spiegelberg, Medex, Epidyn and Gaeltec). Measurement accuracy was best with HanniSet probes. The maximum errors with this transducer were 3 mmHg. Camino and Codman showed similar results. Spiegelberg had slightly larger deviations. With Epidyn and Gaeltec the highest error were noted, up to 10 mmHg in the high pressure range. The 24 h drift was lowest with HanniSet (0.2 mmHg) and Camino (0.8 mmHg). The largest drifts were seen with Medex, Spiegelberg and Gaeltec (1.8 mmHg). Ten day drift was lowest with HanniSet (0.1 mmHg/day) and Codman (0.2 mmHg/day). The highest long-term drifts were found with Epidyn and Gaeltec (1.5 mmHg/day). Drift did not exhibit a linear pattern. After an initial rise in drift during the first 24-72 h, it decreased slowly during the next 7 days. Most ICP probes revealed measurement inaccuracy and drift. These results emphasize the necessity for ongoing evaluations of ICP probes. Therefore, tests for quality assurance are essential to establish a consistent standard of proficiency of ICP transducers.
More Related Videos
05:01A Detailed Protocol for Physiological Parameters Acquisition and Analysis in Neurosurgical Critical Patients
Published on: October 17, 2017
08:55Translaminar Autonomous System Model for the Modulation of Intraocular and Intracranial Pressure in Human Donor Posterior Segments
Published on: April 24, 2020
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Increased Intracranial Pressure l: Introduction
Increased Intracranial Pressure ll: Pathophysiology