Design and validation of a motion stage for in vitro MR experiments
Y Zhou1, T J Carroll, T M Grist
1Departments of Medical Physics and Radiology, University of Wisconsin-Madison, E3/311 Clinical Science Center, 600 Highland Avenue, Madison, Wisconsin 53792-3252, USA. yong.zhou@med.ge.com
Abstract:
A computer-controlled motion stage was designed to facilitate motion studies and simulations of cardiac magnetic resonance (MR) experiments. The stage was constructed to move a phantom of up to 6 kg at velocities up to 15 cm/sec along a two-dimensional trajectory within an MR scanner. The operation of the motion stage was validated and found to be accurate to within 0.08 cm in displacement and 1.5 cm/sec in velocity (root-mean-square error). No image degradation was observed.J. Magn. Reson. Imaging 1999;10:972-977.