Related Experiment Video
Updated: Jan 28, 2026

Author Spotlight: Enhancing Coronary Artery Revascularization
Published on: September 15, 2023
Local haemodynamics of arterial bypass graft anastomoses
C S Rowe1, T K Carpenter, T V How
1Department of Clinical Engineering, University of Liverpool, UK.
Abstract:
One of the main causes of failure of expanded polytetrafluoroethylene (PTFE) bypass grafts used in the lower limbs is the development of myointimal hyperplasia (MIH). Clinical studies show that higher patency rates can be obtained with the use of an autologous vein cuff (the Miller cuff) interposed between the graft and artery. The reasons for the improved performance are still unclear, but preliminary studies suggest that the change in local haemodynamics due to the cuff geometry may be the significant factor rather than the presence of autologous material. If this is the case, then PTFE grafts can be produced with an integral cuff, i.e. a precuffed graft, with similar haemodynamic patterns to that of the Miller cuff. In this paper, two different types of precuffed graft are presented and their flow patterns are compared with those recorded in the Miller cuff and the conventional end-to-side anastomosis. The haemodynamic studies were carried out using optically clear silicone rubber models under simulated in vivo pulsatile flow conditions. Flow structures similar to those observed in the Miller cuff were seen in the precuffed grafts.
Related Concept Videos
Local Attraction
Anastomoses
Anastomoses can be formed at arterial, venous, and lymphatic vessels.
Arterial Anastomosis: These occur between arteries. They...
Arteries and Arterioles
Arteries of Lower Limbs
Local Anesthetics: Pharmacokinetics
Arteries of the Upper Limbs