Analysis of radiation scatter during angiographic procedures: evaluation of a phantom model and a modified radiation
1Department of Radiology, Yamagata University School of Medicine, Yamagata City, Japan. hitoh@med.id.yamagata-u.ac.jp
Purpose:
To study the radiation scattering associated with the digital subtraction angiography (DSA) unit in angiographic procedures and to design an effective radiation protection shield based on these data.
Materials And Methods:
The number of scattered photons was measured at three points relative to the operator's position. Anteroposterior abdominal and lateral cranial fluoroscopy were evaluated. As protective devices, a lead curtain, sliding shields, and a brim-shaped image intensifier (II) hood were designed.
Results:
In abdominal fluoroscopy, radiation was found to scatter to the operator's lower limbs from the underside of the catheter table, to the abdomen from the side of the patient's body, and to the head and neck from the table surface adjacent to the patient. The use of protective devices reduced exposure from 2.89 to 0.058 mR/min for the operator's lower limbs, from 0.987 to 0.069 mR/min for the operator's abdomen, and from 0.696 to 0.139 mR/ min for the operator's head and neck area. With lateral cranial fluoroscopy, radiation was detected to scatter to the operator's lower limbs from the underside of the catheter table, to the abdomen from the patient's temporal area, and to the head and neck from the patient's face. The use of protective devices reduced exposure from 0.248 to 0.010 mR/min for the operator's lower limbs, from 0.129 to 0.010 mR/min for the operator's abdomen, and from 0.162 to 0.018 mR/min for the operator's head and neck area.
Conclusions:
The characteristic directions of scattering to the operator were identified. An effective modified radiation protection system was designed based on this information.
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...