Size-adjustable titanium plate for reconstruction of the sella turcica. Technical note
K Arita1, K Kurisu, A Tominaga
1Department of Neurosurgery, Hiroshima University School of Medicine, Japan. karita@mcai.med.hiroshima-u.ac.jp
Abstract:
A size-adjustable plate constructed of pure titanium is proposed for use in the reconstruction of the sella turcica. The plate is composed of two semicircular pieces that are connected by a hinge located at the top of the plate. Using an applicator, the plate is inserted into the sella turcica in a closed position. The same applicator is then used to open and secure the plate. The titanium causes minimal ferromagnetic artifacts on postoperative magnetic resonance imaging. Preliminary findings indicate a possible clinical use for this plate in the reconstruction of the sella turcica when no suitable piece of bone is available.