Bacteremia associated with tunneled, cuffed hemodialysis catheters

T F Saad1

  • 1Department of Medicine, Division of Nephrology, University of Texas Health Science Center at San Antonio, San Antonio, TX, USA. tfsaad@aol.com

Insights

Catheter-associated bacteremia (CAB) is common in hemodialysis patients. Catheter salvage rates are significantly higher when antibiotic treatment is combined with catheter exchange over a guidewire.

Area of Science:

  • Nephrology
  • Infectious Diseases
  • Medical Devices

Background:

  • Bacteremia is a frequent complication of permanent catheters (PCs) used for hemodialysis (HD) vascular access.
  • The incidence, spectrum of pathogens, and optimal treatment for catheter-associated bacteremia (CAB) require further clarification.

Purpose of the Study:

  • To prospectively monitor infection rates and outcomes in chronic hemodialysis patients with permanent catheters.
  • To evaluate the effectiveness of different treatment strategies for catheter-associated bacteremia, including catheter salvage methods.

Main Methods:

  • Prospective monitoring of 101 chronic hemodialysis patients with permanent catheters over 24 months.
  • Recording data on catheter-days, bacteremia episodes, causative organisms, treatment methods, complications, and outcomes.
  • Comparing outcomes of catheter salvage with antibiotics alone versus antibiotics combined with catheter exchange.

Main Results:

  • An incidence of 5.5 episodes of CAB per 1,000 catheter-days was observed.
  • Gram-positive cocci caused 52.3% of infections, while gram-negative rods accounted for 26.7%.
  • Catheter salvage success rates were significantly higher with antibiotic therapy plus catheter exchange (81.4%) compared to antibiotics alone (36.7%) (P = 0.0005).

Conclusions:

  • Catheter-associated bacteremia is a common complication in hemodialysis units, often involving gram-negative bacteria.
  • Catheter salvage is significantly improved when antibiotic treatment is combined with catheter exchange over a guidewire.