Scatter and veiling glare estimation based on sampled primary intensity.

Y Zhou1, T Mathur, S Molloi

  • 1Department of Radiological Sciences, University of California, Irvine 92697, USA.

Medical Physics
|December 10, 1999
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Adapting TB services during the COVID-19 pandemic in Mumbai, India.

The international journal of tuberculosis and lung disease : the official journal of the International Union against Tuberculosis and Lung Disease·2020
Same author

TU-E-217BCD-09: The Feasibility of the Dual-Dictionary Method for Breast Computed Tomography Based on Photon-Counting Detectors.

Medical physics·2017
Same author

Calibration phantoms for accurate water and lipid density quantification using dual energy mammography.

Physics in medicine and biology·2017
Same author

Energy response calibration of photon-counting detectors using x-ray fluorescence: a feasibility study.

Physics in medicine and biology·2014
Same author

Volumetric lean percentage measurement using dual energy mammography.

Medical physics·2011
Same author

New method to measure coronary velocity and coronary flow reserve.

American journal of physiology. Heart and circulatory physiology·2011
Same journal

Correction to "On the shape of the radiation survival curve in tumor spheroids: The role of oxygen heterogeneity".

Medical physics·2026
Same journal

Multi-view constrained semi-supervised vertebra detection for 3D ultrasound spine volume.

Medical physics·2026
Same journal

Accuracy of quantitative <sup>177</sup>Lu SPECT/CT imaging: A systematic review.

Medical physics·2026
Same journal

Physics-constrained dual-domain network for CBCT reconstruction from orthogonal X-rays in gynecologic radiotherapy.

Medical physics·2026
Same journal

Decomposition-based harmonization for quantitative PET imaging across scanners and radiotracers.

Medical physics·2026
Same journal

Development and evaluation of an in vivo dose-based monitoring system for electron FLASH radiation therapy.

Medical physics·2026
See all related articles

This study introduces a novel scatter measurement technique for videodensitometric iodine quantification, significantly reducing patient x-ray exposure. The new method accurately estimates scatter and veiling glare with minimal radiation dose.

Area of Science:

  • Medical Physics
  • Radiological Imaging
  • Image Analysis

Background:

  • Scatter and veiling glare are major error sources in videodensitometric iodine quantification.
  • Traditional methods using lead strips or discs increase patient x-ray exposure.
  • A need exists for accurate scatter measurement with reduced radiation dose.

Purpose of the Study:

  • To investigate a new scatter measurement technique based on sampled primary intensity.
  • To evaluate the accuracy and efficiency of this technique in digital radiography.
  • To minimize patient x-ray exposure during scatter and veiling glare assessment.

Main Methods:

  • A novel technique using an aperture array in a lead sheet to sample primary x-ray intensity.
  • Calculating scatter-glare intensity by subtracting sampled primary intensity from open field images.

Related Experiment Videos

  • Employing bi-cubic interpolation and digital filtration for pixel-by-pixel scatter-glare estimation.
  • Validation using Lucite step and anthropomorphic chest phantoms.
  • Main Results:

    • The new technique achieved average root-mean-square percentage errors of 8.02% for scatter/veiling glare and 10.01% for primary intensity with bi-cubic interpolation.
    • Digital filtration yielded slightly better results with errors of 7.53% and 8.91%, respectively.
    • The x-ray exposure-area product (EAP) using the aperture array was only 4.38% of the open field EAP.

    Conclusions:

    • The sampled primary intensity technique accurately estimates scatter-glare intensity.
    • This method significantly reduces patient x-ray exposure compared to standard techniques.
    • The findings support the clinical utility of this low-dose scatter measurement approach.