Pseudofracture of an ICD lead
J P Camous1, F Raybaud, J Jourdan
1Medical Department of Cardiology Faculty of Medecine, Nice, France.
A pseudofracture in a defibrillator lead near the implantable cardioverter-defibrillator (ICD) pulse generator was identified. Routine lead impedance measurement during ICD replacement is recommended to detect such issues.
Area of Science:
- Cardiology
- Medical Device Engineering
- Radiology
Background:
- Implantable cardioverter-defibrillators (ICDs) are crucial for managing cardiac arrhythmias.
- Lead integrity is essential for effective ICD function and patient safety.
- Pseudofractures can compromise lead performance and lead to device malfunction.
Purpose of the Study:
- To report a case of proximal pseudofracture in a Medtronic 6936 defibrillating lead.
- To highlight the limitations of conventional X-ray in visualizing critical lead connections.
- To emphasize the importance of lead impedance monitoring during ICD procedures.
Main Methods:
- Radiographic examination (X-ray) was used to identify the lead abnormality.
- The anatomical location and nature of the pseudofracture were described.
- The clinical context of ICD replacement was noted.
Main Results:
- A proximal pseudofracture was detected on X-ray adjacent to the ICD pulse generator.
- The pseudofracture occurred at the multifilar metallic connector site where the stimulating lead joins.
- Conventional abdominal X-rays were insufficient to visualize this specific connection.
Conclusions:
- Proximal lead pseudofractures near the ICD pulse generator can occur.
- Standard X-rays may not reveal critical lead connector issues.
- Routine measurement of lead impedance during ICD replacement is vital for detecting lead integrity problems.
More Related Videos
08:00Reduction of Iatrogenic Atrial Septal Defects with an Anterior and Inferior Transseptal Puncture Site when Operating the Cryoballoon Ablation Catheter
Published on: June 15, 2015
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Related Concept Videos
Flail Chest-I
Flail chest is a severe and potentially life-threatening condition characterized by the fracture of three or more adjacent ribs in multiple places. It is most commonly caused by direct impacts and trauma, such as motor vehicle accidents or injuries from a steering wheel impact. It can also occur due to falls in elderly individuals with osteoporosis, or assaults involving sharp objects.
Pathophysiology
The pathophysiology of flail chest is complex, involving fractures of...
Aneurysm I: Introduction