1Mayo Sleep Disorders Center, Rochester, MN 55905, USA. Hauri.Peter@mayo.edu
Study Objective:
To evaluate the performance of a sleep switch device (held between finger and thumb) in normal sleepers and in insomniacs by comparing its performance to PSG data, wrist actigraphy, and subjective estimates.
Design:
19 insomniacs and 6 normal sleepers each spent one night in the sleep laboratory. Sleep latency and total sleep time were evaluated by the sleep switch device, polysomnography, wrist actigraphy, and subjective estimates.
Setting:
NA.
Patients Or Participants:
NA.
Interventions:
NA.
Results:
The sleep switch device correlated .98 with PSG sleep latency (to the onset of the first 10 minutes of solid sleep, stage 1 or deeper). Mean and standard deviations were almost identical between the two measures. The sleep switch was somewhat less accurate than the wrist actigraph in assessing total sleep time.
Conclusions:
The sleep switch device is a very accurate and inexpensive tool when trying to assess sleep latency in insomniacs and in normals.