Closed-loop control of propofol anaesthesia.

G N Kenny1, H Mantzaridis

  • 1University Department of Anaesthesia, Glasgow Royal Infirmary, UK.

British Journal of Anaesthesia
|January 5, 2000
PubMed
Summary

This study demonstrates an automated closed-loop system for propofol anesthesia using the auditory evoked potential index (AEPindex) to accurately measure anesthetic depth. The system successfully maintained anesthesia with minimal patient movement and satisfactory cardiovascular stability.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Sedation for endoscopy.

Current opinion in anaesthesiology·2006
Same author

Feedback control of anaesthesia.

Current opinion in anaesthesiology·2006
Same author

Current and future applications of target-controlled infusions.

Drugs of today (Barcelona, Spain : 1998)·2003
Same author

Effect site concentrations of remifentanil and pupil response to noxious stimulation.

British journal of anaesthesia·2003
Same author

Effects of the auditory stimuli of an auditory evoked potential system on levels of consciousness, and on the bispectral index.

British journal of anaesthesia·2002
Same author

Auditory evoked potential index predicts the depth of sedation and movement in response to skin incision during sevoflurane anesthesia.

Anesthesiology·2001

Area of Science:

  • Anesthesiology
  • Biomedical Engineering
  • Neuroscience

Background:

  • Accurate monitoring of anesthetic depth is crucial for patient safety.
  • Existing methods for assessing anesthetic depth can be subjective or invasive.
  • Auditory evoked potential index (AEPindex) shows promise as a reliable objective measure.

Purpose of the Study:

  • To develop and validate a closed-loop system for automated propofol anesthesia control.
  • To assess the efficacy of the auditory evoked potential index (AEPindex) as a real-time indicator of anesthetic depth.
  • To evaluate the system's performance in terms of anesthetic control, patient safety, and comfort.

Main Methods:

  • A closed-loop system was designed utilizing the auditory evoked potential index (AEPindex) as the input signal.

Related Experiment Videos

  • A proportional integral (PI) controller was employed to adjust propofol infusion rates based on the AEPindex.
  • The system was tested on 100 spontaneously breathing patients undergoing general anesthesia.
  • Main Results:

    • The AEPindex reliably differentiated between anesthetic states: pre-induction (73.5), surgical anesthesia (37.8), and recovery (89.7).
    • The automated system achieved satisfactory control of anesthesia, with minimal intraoperative awareness or movement.
    • Cardiovascular stability was maintained, and 100% of patients expressed willingness to undergo the same anesthetic in the future.

    Conclusions:

    • Closed-loop automated anesthesia using the AEPindex is a safe and effective method for propofol administration.
    • The AEPindex serves as a valid and reliable measure for real-time depth of anesthesia monitoring.
    • This technology offers potential for improved anesthetic management and patient outcomes.