Improving the positional accuracy of the goniometer on the Philips CM series TEM
J Pulokas1, C Green, N Kisseberth
1Beckman Institute for Advanced Science and Technology, University of Illinois at Urbana-Champaign, 405 North Mathews Avenue, Urbana, Illinois, 61801, USA.
Journal of Structural Biology
|January 14, 2000
Abstract:
We have developed a method to improve the accuracy for absolute relocation of a target specimen using the goniometer on a Philips transmission electron microscope. We have achieved this by characterizing the performance of the Philips compustage, modeling its behavior, and using this model to calculate the goniometer movements required for accurate target relocation. This resulted in a 10-fold improvement in the positioning accuracy of the goniometer.
Related Concept Videos
Common Leveling Mistakes and Errors
A survey team is tasked with determining the elevation difference between points Point A and Point B, separated by uneven terrain. They use a leveling instrument and a leveling rod.Common MistakesMisreading the Rod: During a backsight reading at Point A, the instrumentman observes the rod partially obscured by tall grass. Instead of reading 1.135 m, they mistakenly record 1.735 m due to the misalignment of the crosshair with the wrong graduation. This error adds 0.600 m to all subsequent...
Adjusting a Traverse
In the site survey of a four-sided traverse, internal angles are essential to ensure geometric accuracy. The survey revealed that the sum of the measured internal angles was 359 degrees and 48 minutes, which is 12 minutes less than the expected 360 degrees. This discrepancy signals an error likely arising from measurement inaccuracies during the fieldwork.To rectify this error, the adjustment process involved distributing the 12-minute shortfall equally across the four internal angles. By...