Related Experiment Videos
Development of a microparticle-enhanced nephelometric immunoassay for quantitation of human lysozyme in pleural
M Caballero1, R Ruiz, M Márquez de Prado
1Laboratorio de Bioquímica, Hospital San Millán, Logroño, La Rioja, Spain.
Abstract:
A microparticle-enhanced nephelometric immunoassay, based on polystyrene beads coated with antihuman lysozyme antibody, has been developed for lysozyme quantification in sera and pleural effusions. The standard curve extends from 0.58 mg/l to 18.75 mg/l and no antigen effect was observed. The results showed a good serial precision. The intra-assay precision (n = 20) expressed as CV was between 2.2 and 4.2 in three different concentrations. The inter-assay precision, with different calibration curves (n = 12) was between 6.4 and 7.1. The analytical assay showed a sufficient linearity (r > 0.999). There were no interferences either with haemoglobin (up to 4 g/l), lipids (up to 0.5%, expressed as 1% Lipofundina content), or bilirubin (up to 5 mg/dl). The analytical sensitivity was lower than 0.6 mg/l. The correlation with a Micrococcus lysodeikticus turbidimetric assay showed a correlation coefficient of 0.915. We have studied 92 patients with pleural effusion. In each case, pleural fluid adenosine deaminase activity and pleural fluid to plasma lysozyme ratio were determined. The lysozyme ratio showed similar clinical sensitivity and specificity as to adenosine deaminase.