Admixture with whole blood does not explain false-negative urease tests.
1Department of Medicine, Sinai Hospital, Wayne State University, Detroit, Michigan 48235, USA.
Journal of Clinical Gastroenterology
|January 15, 2000
Summary
Whole blood enhances rapid urease tests for Helicobacter pylori in vitro. This study found that blood admixture does not cause false negatives, suggesting other factors affect test sensitivity during gastrointestinal bleeding.
Area of Science:
- Gastroenterology
- Clinical Microbiology
- Diagnostic Testing
Background:
- Rapid urease tests (RUTs) are commonly used for Helicobacter pylori detection.
- Reduced RUT sensitivity is observed in patients with active upper gastrointestinal bleeding (UGIB).
- The in vitro effect of whole blood on RUT performance remains unclear.
Purpose of the Study:
- To investigate the in vitro impact of whole blood on the sensitivity and accuracy of rapid urease testing for H. pylori.
- To determine if blood admixture contributes to false-negative RUT results in UGIB.
Main Methods:
- Urease solution was mixed with saline and heparinized whole blood (H. pylori antibody-positive and -negative).
- Mixtures were serially diluted in saline and/or whole blood.
- Admixtures were tested using three different rapid urease kits.
Main Results:
- Whole blood admixture, particularly in H. pylori-seropositive blood, enhanced RUT performance up to fourfold compared to saline alone.
- No false-negative results were observed in any tested kit due to whole blood admixture.
- Both H. pylori antibody-positive and -negative whole blood accelerated the urease reaction in vitro without adverse impact.
Conclusions:
- Whole blood accelerates the urease reaction in vitro, potentially improving RUT performance.
- Admixture with whole blood does not appear to cause false-negative RUT results.
- The reduced sensitivity of RUT in active UGIB is likely due to factors other than whole blood interference.