Anomalous dose assessment associated with film badge storage in a controlled area
1Radiology Department, Aultman Hospital, Northeastern Ohio Universities College of Medicine, Canton 44710-1799, USA. wrhedric@neoucom.edu
Abstract:
Higher than expected film badge readings were obtained for radiologic technologists operating computed tomography scanners. The cause of the erroneous dose assignments was determined to be improper storage of the film packet in the console area. Since the filter holder was absent during exposure, the overestimation of photon energy resulted in an incorrect conversion from optical density to dose. Film badges and particularly the film packet should not be stored in controlled areas where the radiation levels are expected to be above background.