A Methodological Approach to Non-invasive Assessments of Vascular Function and Morphology
Published on: February 7, 2015
Detection of vascular haemodynamics through a high-speed velocity profiler
1Electronic Engineering Department, University of Florence, via S. Marta 3, 50139, Florence, Italy. ultra@infgil.ing.unifi.it
Objective:
This paper aims at demonstrating that ultrasound Doppler multigate spectral analysis performed with advanced equipment may provide detailed and significant haemodynamic information.
Methods:
A novel multigate system was recently introduced and shown capable of performing real-time spectral analysis of Doppler data from 64 resolution cells located at different depths from the transducer. The system extends the typical capabilities of conventional Pulsed Wave (PW) equipment by displaying the full spectral content of Doppler signals over an ultrasound scan line rather than in a single resolution cell. In cases where it is appropriate to display the available information in a simpler form, parameters such as the maximum frequency can be extracted from each spectrum, by using conventional or advanced image processing methods.
Results:
In-vitro experiments show that the multigate system can perform velocity measurements with good accuracy and precision. Examples of in vivo profiles detected from carotid, femoral and radial arteries are presented. In particular, the first results obtained from the aorta are shown.
Conclusions:
Blood flow behavior can be accurately investigated using a real-time multigate system which extends Doppler spectral analysis to a whole scan line.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...