Related Experiment Videos
Evaluation of the RapID Yeast Plus System for the identification of yeast
J Moghaddas1, A L Truant, C Jordan
1Temple University Hospital and School of Medicine, Philadelphia, PA 19140, USA.
Diagnostic Microbiology and Infectious Disease
|February 11, 2000
Abstract:
We evaluated the RapID Yeast Plus System using 117 fresh and frozen clinical yeast isolates. The Uni-Yeast-Tek System was used to establish the correct identification. The Vitek System was used as the arbiter for any discrepant results, along with morphology. Of 117 isolates tested, the RapID Yeast Plus System identified 96.6% correctly. The RapID Yeast Plus System is an accurate and reliable alternative to other commonly used yeast identification systems.