Nonaqueous capillary electrophoresis equipped with amperometric detection for analysis of chlorinated phenolic
J H Luong1, A Hilmi, A L Nguyen
1Biotechnology Research Institute, National Research Council Canada, Montreal, Quebec.
Abstract:
Nonaqueous capillary electrophoresis (NACE) equipped with amperometric detection has been developed for separation and detection of an 11-member model mixture of chlorinated phenolic compounds. With triacetyl-beta-cyclodextrin (TACD) as a novel selectivity selector, acetonitrile proved to be an excellent solvent for this water-insoluble cyclodextrin derivative. Resolution of the analytes was achieved by using an optimized acetonitrile medium consisting of 500 mM acetic acid, 10 mM sodium acetate, 12 mM TACD and 50 mM tetrabutylammonium perchlorate. Separation of analytes was attributed to differential electrostatic and/or inductive interactions of the analytes with the TACD/TBA+ complex and charged tetrabutylammonium phases. A simple end-column amperometric detector (Pt vs. Ag/AgCl, poised at +1.6 V) in conjunction with NACE was used to analyze chlorophenols. Amperometric detection of such target compounds in acetonitrile-based media offers high sensitivity and alleviates electrode fouling compared to aqueous buffers. The detection limits obtained, ranging from 30 nM to 500 nM, are 3-8-fold lower than those obtained with aqueous buffers.
More Related Videos
10:13Using Capillary Electrophoresis to Quantify Organic Acids from Plant Tissue: A Test Case Examining Coffea arabica Seeds
Published on: November 12, 2016
04:39A Modified QuEChERS-HPLC Method for Detection of Polycyclic Aromatic Hydrocarbons in Zebrafish Embryos Exposed to Fine Particulate Matter
Published on: June 13, 2025
Related Concept Videos
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
High-Performance Liquid Chromatography: Types of Detectors
Ion-Exchange Chromatography
Electrophoresis: Overview
There...
Capillary Electrophoresis: Instrumentation
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...