Accuracy of carboxyhemoglobin dilution method for the measurement of circulating blood volume

S Ohki1, F Kunimoto, Y Isa

  • 1Second Department of Surgery, Gunma University School of Medicine, Maebashi, Japan. ohkisato@akagi.sb.gunma-u.ac.jp

Insights

The carbon monoxide-labeled hemoglobin (CO-Hb) dilution method accurately measures circulating blood volume (BVc) in ICU patients, offering a safe and repeatable alternative to radioactive methods.

Area of Science:

  • Critical Care Medicine
  • Cardiovascular Surgery
  • Diagnostic Methods

Background:

  • Accurate management of circulating blood volume (BVc) is vital for intensive care unit (ICU) patients.
  • Existing methods for BVc measurement may have limitations.

Purpose of the Study:

  • To evaluate the accuracy and precision of the carbon monoxide-labeled hemoglobin (CO-Hb) dilution method (CO method) for BVc measurement.
  • To compare the CO method against the established 51Cr-labeled erythrocyte dilution method (51Cr method).

Main Methods:

  • A prospective study involving 18 patients undergoing coronary artery bypass grafting (CABG) with cardiopulmonary bypass (CPB).
  • BVc was measured using both CO and 51Cr methods 24 hours post-ICU admission.
  • Paired data analysis and percentage error calculations were used to assess agreement.

Main Results:

  • The CO method demonstrated small mean differences and standard deviations compared to the 51Cr method (-70.2 +/-184.8 mL).
  • Low percentage errors (-0.49+/-1.29%) and a high correlation coefficient (r = 0.97) confirmed accuracy and precision.
  • The CO method showed excellent agreement with the 51Cr method.

Conclusions:

  • The CO method provides accurate BVc measurements comparable to the 51Cr method.
  • The CO method is simple, repeatable, and safe, avoiding radioactivity exposure in the ICU setting.
  • This offers a valuable, non-radioactive option for BVc monitoring in critical care.