Related Experiment Videos
Visualization of viscous fingering in high-performance liquid chromatographic columns. Influence of the header design
R A Shalliker1, B S Broyles, G Guiochon
1Department of Chemistry, The University of Tennessee, Knoxville 37996-1600, USA.
Abstract:
Using an on-column visualization technique, band profiles of solutes migrating along an HPLC column were studied. The study showed that, under conditions where viscous fingering is prevalent, the design of the inlet header has little influence on the outcome of the viscous fingers. Two types of headers were studied. The first contained a small diameter inlet frit, which localized the majority of the sample in or near the central region of the column. The second header contained a wide frit and produced a more uniform radial distribution of the sample. In both cases, the extent of viscous fingering was essentially the same.