Enantiomer separation of alpha-ionone using gas chromatography with cyclodextrin derivatives as chiral stationary
F Quattrini1, G Biressi, M Juza
1Laboratorium für Technische Chemie, ETH Zürich, Switzerland.
Abstract:
The gas chromatographic enantiomer separation of alpha-ionone was studied with three different chiral stationary phases using as chiral selectors: (1) heptakis(2,3,6-tri-O-methyl)-beta-cyclodextrin, dissolved in polysiloxane PS-086, (2) octakis(2,6-di-O-pentyl-3-O-trifluoroacetyl)-gamma-cyclodextrin and (3) octakis(2,6-di-O-pentyl-3-O-butanoyl)-gamma-cyclodextrin, both dissolved in polysiloxane SE-54. The influence of the concentration of the chiral selector in the polysiloxane, coated on Chromosorb P AW-DMCS 80-100 mesh, is described and discussed, as well as the effect of Chromosorb loading. The feasibility of the preparative gas chromatographic separation of the enantiomers of alpha-ionone is considered; in order to provide a term of comparison, the estimated performances are compared with those achieved in the separation of the enantiomers of the inhalation anaesthetic enflurane.
Related Concept Videos
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Ion-Exchange Chromatography
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
Chromatography: Introduction
The phase in which the compounds linger or on which the compounds adsorb is called the stationary phase, whereas the mobile phase is the solvent that carries the solutes to be analyzed. In traditional column chromatography, the mixture flows through the stationary phase, and the compounds partition between the stationary and mobile phases...
Analyte Adsorption and Distribution