Related Experiment Video
Updated: Aug 4, 2026

Monitoring the Wall Mechanics During Stent Deployment in a Vessel
Published on: May 8, 2012
A technique to prevent newly implanted stent displacement during subsequent catheter and sheath manipulation
M R Recto1, F F Ing, R G Grifka
1Lillie Frank Abercrombie Section of Pediatric Cardiology, Texas Children's Hospital, Baylor College of Medicine, Houston, Texas, USA. mitch.recto@luisville.edu
Abstract:
We describe a novel technique to prevent the displacement or migration of a newly implanted stent as a consequence of any subsequent catheter and sheath manipulation during the same catheterization procedure. The technique involves reinflation of the dilation balloon within the stent immediately after implant followed by advancing the long delivery sheath carefully over the balloon as the balloon is slowly deflated within the stent. The technique was used successfully in 78 stents in 30 patients without stent dislodgment or migration.