Related Experiment Videos
Positioning of vagal nerve stimulators: technical note
Surgical Neurology
|March 14, 2000
Summary
Repositioning the vagal nerve stimulator pulse generator during MRI scans prevents deactivation. This ensures continuous treatment for epilepsy patients and allows for necessary functional MRI (fMRI) evaluations.
Area of Science:
- Neurology
- Biomedical Engineering
Background:
- Vagal nerve stimulation (VNS) is a key treatment for intractable seizures.
- Patients with VNS often require brain MRIs to monitor neurological conditions.
- Current implant techniques risk VNS deactivation during MRI scans.
Purpose of the Study:
- To identify an MRI-compatible positioning technique for VNS pulse generators.
- To prevent unintended deactivation of VNS devices during magnetic resonance imaging.
Main Methods:
- Assessed a VNS pulse generator model using an oscilloscope and LED in an MRI environment.
- Simulated device performance in various positions.
- Studied patients with implanted VNS devices to confirm stimulator activation.
Main Results:
- Placing the pulse generator with electrode inputs parallel to the body's long axis prevented deactivation in the MRI's magnetic field.
- This positioning maintained device function during supine scans.
Conclusions:
- Altering the VNS pulse generator's implant orientation is a simple solution to prevent MRI-induced deactivation.
- This allows for essential MRI and functional MRI (fMRI) scans without interrupting VNS therapy.