Ligating clips for three-dimensional MR angiography at 1.5 T: in vitro evaluation
Related Experiment Video
Updated: Oct 17, 2026

Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
D Weishaupt1, H H Quick, D Nanz
1Institute of Diagnostic Radiology, University Hospital Zürich, Switzerland.
Abstract:
Artifact size on three-dimensional (3D) magnetic resonance (MR) angiograms and safety of various vascular clips (15 titanium and three absorbable polydioxanone clips) were assessed. All evaluated clips were completely safe. Biodegradable clips rendered no artifacts; titanium clips were associated with susceptibility effects. Artifact size was dependent on clip size, clip orientation, echo time, and degree of k-space coverage. In the presence of titanium vascular clips, fast 3D MR angiography should be performed with the shortest echo time and full k-space coverage.