Related Experiment Video
Updated: Oct 17, 2026

Automated 90Sr Separation and Preconcentration in a Lab-on-Valve System at Ppq Level
Published on: June 6, 2018
Influence of radioactive impurities on SIR measurements
1Bureau International des Poids et Mesures (BIPM), Sevres, France.
Abstract:
The correction usually applied to an SIR measurement in the presence of a gamma-ray emitting impurity is presented and explained. The method of calibration of a Ge(Li) spectrometer is briefly described. It is used to look for impurities when SIR measurements show inconsistencies. An example is presented where a 103Ru impurity was discovered in a 99Mo solution. The correction applied compensates for the decrease of around 0.2% per day observed in the SIR measurements of this 99Mo ampoule.
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Isotopes and Radioisotopes
An isotope containing more...
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Biological Effects of Radiation
Radioactive Decay and Radiometric Dating
Atomic Emission Spectroscopy: Interference