The Use of Mixed Reality in Custom-Made Revision Hip Arthroplasty: A First Case Report
Published on: August 4, 2022
[Complete dislocation of the polyethylene inlay after internal hemipelvectomy and individually fitted tumor
T K Lichtinger1, M Lederer, R T Müller
1Orthopädische Universitätsklinik Essen.
Abstract:
We report about an exceptional complication after internal hemipelvectomy and replacement of the defect with a custom-made endoprosthesis. A complete luxation of the PE-inlay out of the metal cup occurred. Radiologically we assumed a luxation of the femoral head in dorsal direction. Revision after a failed attempt of closed reposition showed a complete luxation of the inlay, which was caused by deficient fixation in the acetabular component. The absence of a contrast wire in the PE-inlay delayed the right diagnosis and made it difficult to find the dislocated inlay in the large wound.