Frontline immunochromatographic device for on-site urine testing of amphetamines: laboratory validation using
1Department of Clinical Pharmacology, Karolinska Hospital, Stockholm, Sweden. olof.beck@mb.ks.se
Annals of Clinical Biochemistry
|March 29, 2000
Summary
The Frontline test accurately screens for amphetamines and methamphetamines in urine, showing high sensitivity and specificity. While generally reliable, some results between 300-1000 ng/mL may be affected by antibody enantioselectivity.
Area of Science:
- Clinical Chemistry
- Forensic Toxicology
- Analytical Chemistry
Background:
- Rapid screening tests are crucial for detecting amphetamines and methamphetamines in clinical and forensic settings.
- Immunochromatographic devices offer a quick, visual method for drug testing.
- Accurate performance evaluation of new screening devices is essential.
Purpose of the Study:
- To evaluate the performance of a new immunochromatographic test device (Frontline) for detecting amphetamines and methamphetamines.
- To compare the Frontline device results with established immunoassays and gas chromatography methods.
- To assess the device's sensitivity, specificity, and agreement across different drug concentration levels.
Main Methods:
- Utilized 658 authentic clinical and forensic urine specimens.
- Compared the Frontline test results against established immunoassays.
- Investigated discordant results using gas chromatography-mass spectrometry (GC-MS) or gas chromatography-flame ionization detection (GC-FID).
Main Results:
- The Frontline device demonstrated a sensitivity of 93% and a specificity of 98%.
- High agreement was observed at amphetamine concentrations below 150 ng/mL and above 1000 ng/mL.
- A subset of specimens with concentrations between 300-1000 ng/mL yielded false-negative results, potentially due to antibody enantioselectivity towards d-amphetamine.
Conclusions:
- The Frontline test device provides adequate performance for presumptive screening of amphetamines and methamphetamines.
- The observed discrepancies in the mid-range concentration zone warrant consideration in interpretation.
- Further investigation into the enantioselectivity of antibodies is recommended for optimizing test accuracy.