Related Experiment Video
Updated: Oct 17, 2026

Imaging In-Stent Restenosis: An Inexpensive, Reliable, and Rapid Preclinical Model
Published on: September 14, 2009
MR-guided angioplasty of renal artery stenosis in a pig model: a feasibility study
1Department of Radiology, Medical Physics, and Pathology, University of Wisconsin-Madison Medical School, USA. raomary@facstaff.wisc.edu
Purpose:
To test the hypothesis that magnetic resonance (MR) imaging can guide the percutaneous treatment of renal artery stenosis in a pig model.
Materials And Methods:
Ameroid constrictors were surgically placed around six renal arteries in four pigs. After 30-36 days, all stenoses were documented by conventional x-ray aortograms. MR-guided renal angioplasty was attempted for three stenoses. For these pigs, MR angiography was performed with use of contrast-enhanced three-dimensional (3D) techniques. The authors visualized catheters by filling them with dilute 4% gadolinium and imaging with two-dimensional (2D) and 3D MR fast spoiled gradient recalled echo techniques. Under MR guidance, the authors advanced a selective catheter into the affected renal artery and crossed the stenosis with a nitinol guide wire. Angioplasty was performed with a balloon catheter filled with dilute gadolinium. Stenosis and luminal diameter measurements were compared before and after angioplasty.
Results:
After ameroid constrictor placement, four significant stenoses, one mild stenosis, and one occlusion developed. Under MR guidance, the authors achieved technical success in performing three of three (100%) attempted dilations. After MR-guided angioplasty, the mean reduction in stenosis was 35% and the mean increase in luminal diameter was 1.6 mm.
Conclusion:
Use of MR guidance for the angioplasty of renal artery stenosis in pigs is feasible.