Related Experiment Videos
Electromechanical characterization of acute experimental myocardial infarction
R Kornowski1, M K Hong, A Shiran
1Cardiology Research Foundation, Department of Cardiology and Medlantic Research Institute, Washington Hospital Center, 110 Irving Street, N.W., 4B-1, Washington, D.C. 20010, USA. rxk3@mhg.edu
Objective:
A new cardiac mapping system combines harmless magnetic field energy and tip-deflecting catheters (equipped with location sensors) to obtain real-time 3-dimensional electromechanical maps of the left ventricle endocardial surface without using x-ray fluoroscopy. This experimental study assessed electromechanical changes during acute coronary occlusion and reperfusion in a canine model.
Methods:
Group 1 (n = 10) underwent coronary occlusion for 45 minutes followed by reperfusion (n = 6) and group 2 (n = 11) underwent coronary occlusion for 90 minutes. Endocardial peak-to-peak voltage amplitudes and local endocardial shortening values were measured in ischemic and non-ischemic zones at baseline, following coronary occlusion and reperfusion.
Results:
In ischemic zones, local shortening was significantly reduced during coronary occlusion compared to baseline (Group 1: 4.7 +/- 2.0% at 45 minutes vs. 15.5 +/- 3.4%, p < 0.001, 6.2 +/- 2.1% at 90 minutes vs. 15.5 +/- 3.4%, p < 0.001; Group 2: 5.0 +/- 2.9% at 90 minutes vs. 13.9 +/- 3.3%, p = 0.007). Coronary occlusion caused a significant reduction in voltage potentials in the ischemic area (unipolar voltage at 45 minutes: 32.2 +/- 7.3 mV vs. 36.2 +/- 8.5 mV at baseline, p = 0.03; unipolar voltage at 90 minutes: 30.5 +/- 11.3 mV vs. 38.3 +/- 14.2 mV, p = 0.003; bipolar voltage at 45 minutes: 7.6 +/- 5.5 mV vs. 10.1 +/- 6.0 mV, p < 0.04; bipolar voltage at 90 minutes: 7.6 +/- 4.4 mV vs. 9.8 +/- 6.2 mV, p < 0.02). Voltage amplitudes were no longer reduced during reperfusion (unipolar voltage: 34.3 +/- 10.5 mV vs. 36.2 +/- 8.5 mV, p = 0.26; bipolar voltage: 9.1 +/- 4.5 mV vs. 10.1 +/- 6.0 mV at baseline, p = 0.37), or in non-ischemic regions during either coronary occlusion or reperfusion.
Conclusions:
Electromechanical mapping study provides unique insights into acute myocardial infarction and stunning by detection and localization of early electromechanical changes during coronary occlusion and/or reperfusion.