Related Experiment Video
Updated: Jun 28, 2026

Remote Magnetic Navigation for Accurate, Real-time Catheter Positioning and Ablation in Cardiac Electrophysiology Procedures
Published on: April 21, 2013
Nonfluoroscopic three-dimensional mapping for arrhythmia ablation: tool or toy?
A Khongphatthanayothin1, E Kosar, K Nademanee
1Clinical Electrophysiology and Pacing Center, Department of Internal Medicine, University of Southern California School of Medicine, Los Angeles 90033, USA.
The CARTO system, a nonfluoroscopic 3D mapping technique, significantly reduces fluoroscopy time for cardiac ablation procedures. This advanced method achieves excellent outcomes comparable to conventional fluoroscopy, without increasing procedure duration.
Area of Science:
- Electrophysiology
- Medical Technology
- Cardiac Ablation
Background:
- Conventional fluoroscopy-guided cardiac ablation poses risks of imprecise catheter positioning and prolonged radiation exposure.
- The development of nonfluoroscopic 3D mapping systems offers a potential alternative to improve safety and efficiency.
Purpose of the Study:
- To evaluate the efficacy and safety of the CARTO nonfluoroscopic 3D electroanatomic mapping system.
- To compare CARTO-guided ablation outcomes with conventional fluoroscopy-guided ablation.
Main Methods:
- A comparative study involving 88 arrhythmia ablations using CARTO and 100 ablations using conventional fluoroscopy.
- Ablations included AV nodal reentrant tachycardia, atrial tachycardia/flutter, ventricular tachycardia, and bypass tract tachycardia.
- Outcomes assessed included success rates, complications, fluoroscopy time, and procedure time.
Main Results:
- Excellent and comparable ablation outcomes were achieved with both CARTO and conventional techniques across all arrhythmia types.
- CARTO significantly reduced fluoroscopy times for all studied arrhythmias (e.g., AVNRT: 10+/-7 min vs. 27+/-15 min).
- Procedure times were similar, except for bypass tract tachycardia, which was shorter with CARTO (4+/-1.3 hrs vs. 5.5+/-2.5 hrs).
Conclusions:
- The CARTO electroanatomic 3D mapping technique effectively reduces fluoroscopy time during cardiac ablation.
- This advanced mapping system provides excellent clinical outcomes without compromising procedure efficiency.
More Related Videos
10:46Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019