Four multidetector-row helical CT: image quality and volume coverage speed
Purpose:
To evaluate the imaging performance characteristics of four multidetector-row helical computed tomography (CT) and assess improvement in volume coverage speed over that with single multidetector-row helical CT.
Materials And Methods:
The section-sensitivity profile and image noise of a four multidetector-row scanner were measured with phantom scans and compared with predictions from theoretic models. Nominal section thickness ranged from 1.25 to 10.00 mm, beam collimation from 1.25 to 5.00 mm, and table speed from 3.75 to 30.00 mm per rotation. Image artifacts with four and single multidetector-row helical CT were compared in both a phantom study and a subjective rating analysis of clinical images.
Results:
Compared with single multidetector-row helical CT, the volume coverage speed of four multidetector-row helical CT (range, 3.75-30.00 mm per rotation) is at least twice as fast as that with single multidetector-row helical CT (1.0-10.0 mm per rotation) with fully comparable image quality or, in many cases, three times as fast with diagnostically comparable image quality.
Conclusion:
Compared with single multidetector-row helical CT, four multidetector-row helical CT provides a two- to threefold improvement in volume coverage speed with comparable diagnostic image quality.
More Related Videos
11:49Using High Resolution Computed Tomography to Visualize the Three Dimensional Structure and Function of Plant Vasculature
Published on: April 5, 2013
09:21Human Brown Adipose Tissue Depots Automatically Segmented by Positron Emission Tomography/Computed Tomography and Registered Magnetic Resonance Images
Published on: February 18, 2015
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Imaging Studies IV: Magnetic Resonance Imaging