[Optimizing temporal resolution in CT with retrospective ECG gating].

J M Boese1, M L Bahner, J Albers

  • 1Abteilung Onkologische Diagnostik und Therapie, Deutsches Krebsforschungszentrum Heidelberg. j.boese@dkfz.de

Der Radiologe
|April 12, 2000
PubMed
Summary

New algorithms for time-optimized retrospective cardiac gating significantly improve temporal resolution in spiral CT cardiac imaging. This advancement enables clearer systolic and diastolic images, aiding in the diagnosis of heart conditions.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Free propagation phase-contrast breast CT provides higher image quality than cone-beam breast-CT at low radiation doses: a feasibility study on human mastectomies.

Scientific reports·2019
Same author

Chest wall pain following lung stereotactic body radiation therapy using 48Gy in three fractions: A search for predictors.

Cancer radiotherapie : journal de la Societe francaise de radiotherapie oncologique·2019
Same author

X-ray-Based 3D Virtual Histology-Adding the Next Dimension to Histological Analysis.

Molecular imaging and biology·2018
Same author

Effects of repetitive prolonged breath-hold in elite divers on myocardial fibrosis and cerebral morphology.

European journal of radiology·2018
Same author

[Erratum to: Retention of contrast media in the history of radiology. Sequelae of the former thorotrast use and new challenges.]

Der Radiologe·2017
Same author

[Retention of contrast media in the history of radiology : Sequelae of the former use of thorotrast and new challenges].

Der Radiologe·2016

Area of Science:

  • Medical Imaging
  • Cardiovascular Technology
  • Computed Tomography

Background:

  • Conventional spiral CT cardiac imaging is limited by long acquisition times (250-500 ms) using standard ECG synchronization.
  • This limitation restricts image acquisition to the diastolic phase for patients with moderate heart rates.

Purpose of the Study:

  • To present and investigate algorithms for time-optimized retrospective cardiac gating.
  • To improve temporal resolution in spiral CT cardiac imaging.

Main Methods:

  • Algorithms utilize data from multiple gantry rotations for image reconstruction.
  • Methods applicable to multi-scans at fixed slice positions and overlapping spiral scans.
  • Temporal resolution quantified via computer simulations and validated with experimental data.

Related Experiment Videos

Main Results:

  • Spiral scanning on sub-second CT scanners achieved significantly higher temporal resolutions.
  • A temporal resolution of 170 ms yielded systolic images with minimal motion artifacts.
  • Multi-scans demonstrated temporal resolutions up to 70 ms, enabling visualization of ventricle wall movement throughout the cardiac cycle.

Conclusions:

  • Time-optimized retrospective cardiac gating expands the capabilities of conventional spiral CT for cardiac imaging.
  • The method is transferable to multi-slice scanners, offering reduced scan times for clinical use.
  • Potential applications include determining cardiac functional parameters (e.g., ejection fraction) and detecting ventricle wall motion disorders.