Related Experiment Video
Updated: Aug 19, 2026

A Novel Technique for Raman Analysis of Highly Radioactive Samples Using Any Standard Micro-Raman Spectrometer
Published on: April 12, 2017
Radioactive contamination of packing materials from a xenon-133 shipment
1Radiation Safety Office and Nuclear Medicine Service, Department of Veterans Affairs Medical Center, Lexington, Kentucky 40511-1093, USA. hackett.michael_t@lexington.med.va.gov
Objective:
We report on radioactive contamination of packing materials from a 133Xe shipment.
Methods:
A 2-vial 133Xe shipment was monitored using a survey meter before opening. Both vials were immediately assayed in a dose calibrator. The packing materials were monitored and contamination was detected.
Results:
The maximum surface reading of the shipment was 7.0 microSv/h. This was higher than previous shipments (1.1 +/- 0.3 microSv/h). One vial was 544 MBq while the other vial was only 474 MBq. Previous shipments were 565 +/- 13 MBq/vial. Monitoring and imaging revealed 133Xe contamination within the packing materials. Xenon-133 escaped from the packing materials over time. The lower activity vial continued to leak 133Xe over time.
Conclusion:
Careful monitoring of 133Xe shipments before and after opening along with assaying vials on receipt can indicate vial leakage and radioactive contamination so steps can be taken to minimize radiation exposure to the staff.
Related Concept Videos
Types of Radioactivity
Alpha (α) decay is the emission of an α particle from the nucleus. For example, polonium-210 undergoes α decay:
Nuclear Stability
To hold positively charged protons together in the...
Radioactive Decay and Radiometric Dating
Nuclear Power
Nuclear Fuels
Nuclear fuel consists of a fissile isotope, such as uranium-235, which must be present in sufficient quantity to provide a...
Nuclear Transmutation
Biological Effects of Radiation