Related Experiment Video
Updated: Oct 16, 2026

Investigation into Deep Breathing through Measurement of Ventilatory Parameters and Observation of Breathing Patterns
Published on: September 16, 2019
The effects of passive humidifier dead space on respiratory variables in paralyzed and spontaneously breathing
R S Campbell1, K Davis, J A Johannigman
1Department of Surgery, University of Cincinnati, Ohio 45267-0558, USA. Robert.Campbell@UC.edu
Background:
Passive humidifiers have gained acceptance in the intensive care unit because of their low cost, simple operation, and elimination of condensate from the breathing circuit. However, the additional dead space of these devices may adversely affect respiratory function in certain patients. This study evaluates the effects of passive humidifier dead space on respiratory function.
Methods:
Two groups of patients were studied. The first group consisted of patients recovering from acute lung injury and breathing spontaneously on pressure support ventilation. The second group consisted of patients who were receiving controlled mechanical ventilation and were chemically paralyzed following operative procedures. All patients used 3 humidification devices in random order for one hour each. The devices were a heated humidifier (HH), a hygroscopic heat and moisture exchanger (HHME) with a dead space of 28 mL, and a heat and moisture exchanger (HME) with a dead space of 90 mL. During each measurement period the following were recorded: tidal volume, minute volume, respiratory frequency, oxygen consumption, carbon dioxide production, ratio of dead space volume to tidal volume (VD/VT), and blood gases. In the second group, intrinsic positive end-expiratory pressure was also measured.
Results:
Addition of either of the passive humidifiers was associated with increased VD/VT. In spontaneously breathing patients, VD/VT increased from 59 +/- 13 (HH) to 62 +/- 13 (HHME) to 68 +/- 11% (HME) (p < 0.05). In these patients, constant alveolar ventilation was maintained as a result of increased respiratory frequency, from 22.1 +/- 6.6 breaths/min (HH) to 24.5 +/- 6.9 breaths/min (HHME) to 27.7 +/- 7.4 breaths/min (HME) (p < 0.05), and increased minute volume, from 9.1 +/- 3.5 L/min (HH) to 9.9 +/- 3.6 L/min (HHME) to 11.7 +/- 4.2 L/min (HME) (p < 0.05). There were no changes in blood gases or carbon dioxide production. In the paralyzed patient group, VD/VT increased from 54 +/- 12% (HH) to 56 +/- 10% (HHME) to 59 +/- 11% (HME) (p < 0.05) and arterial partial pressure of carbon dioxide (PaCO2) increased from 43.2 +/- 8.5 mm Hg (HH) to 43.9 +/- 8.7 mm Hg (HHME) to 46.8 +/- 11 mm Hg (HME) (p < 0.05). There were no changes in respiratory frequency, tidal volume, minute volume, carbon dioxide production, or intrinsic positive end-expiratory pressure.
Discussion:
These findings suggest that use of passive humidifiers with increased dead space is associated with increased VD/VT. In spontaneously breathing patients this is associated with an increase in respiratory rate and minute volume to maintain constant alveolar ventilation. In paralyzed patients this is associated with a small but statistically significant increase in PaCO2.
Conclusion:
Clinicians should be aware that each type of passive humidifier has inherent dead space characteristics. Passive humidifiers with high dead space may negatively impact the respiratory function of spontaneously breathing patients or carbon dioxide retention in paralyzed patients. When choosing a passive humidifier, the device with the smallest dead space, but which meets the desired moisture output requirements, should be selected.
More Related Videos
08:25Combining Volumetric Capnography And Barometric Plethysmography To Measure The Lung Structure-function Relationship
Published on: January 8, 2019
09:37Evaluation of Respiratory Muscle Activation Using Respiratory Motor Control Assessment (RMCA) in Individuals with Chronic Spinal Cord Injury
Published on: July 19, 2013
Related Concept Videos
Alterations in Respiration II
In Biot's breathing, the respiratory rate and depth are irregular, alternating between periods of deep gasping and apnea. Common causes include...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Physiological Control of Respiration
Breathing, a seemingly passive process, is regulated by the respiratory center in the brainstem. This center coordinates the involuntary control of respirations, which means it occurs without conscious effort, ensuring a smooth and uninterrupted pattern.
Regulation of Ventilation
The body maintains ventilation by monitoring levels of carbon dioxide (CO2), oxygen (O2), and hydrogen ion concentration (pH) in the arterial blood. Among these factors, the level of CO2 plays a crucial...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Physical Assessment of the Respiratory Tract II: Inspection
Chest Configuration
The chest configuration can...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...