The screening of the central visual field
Summary
The Armaly technique for glaucoma screening showed the Auto-Plot tangent screen had the lowest false positive rate in normal eyes. The Friedmann analyser was easiest to use but less specific for early glaucomatous field defects.
Area of Science:
- Ophthalmology
- Visual Science
- Medical Technology
Background:
- Glaucomatous field defects require accurate and efficient screening methods.
- The Armaly technique is a specific approach to perimetric screening for visual field loss.
Purpose of the Study:
- To evaluate the effectiveness of the Armaly technique using different perimetric instruments.
- To compare the performance of the Goldmann perimeter, Auto-Plot tangent screen, and Friedmann central field analyser in screening for glaucomatous field defects.
Main Methods:
- The Armaly technique was applied to 100 normal eyes and 38 eyes with known central field defects.
- Screening results were obtained using the Goldmann perimeter and Auto-Plot tangent screen.
- These results were compared against those from the Friedmann central field analyser.
Main Results:
- The Auto-Plot tangent screen demonstrated the lowest initial false positive rate in normal eyes.
- The Friedmann analyser exhibited the highest initial false positive rate.
- The Friedmann analyser showed lower specificity for early central field changes compared to the Goldmann perimeter and Auto-Plot tangent screen.
Conclusions:
- The Auto-Plot tangent screen is a more specific tool for Armaly's technique screening than the Friedmann analyser.
- Despite lower specificity, the Friedmann analyser is the easiest instrument for mass screening perimetry due to its simplicity.