Long-term outcome of patients with multiple (> or = 3) noninfected transvenous leads: a clinical and
C C de Cock1, M Vinkers, L C Van Campe
1University Hospital VU, Amsterdam, The Netherlands. cardiol@azvu.nl
Insights
Patients with multiple transvenous leads (≥3) face no increased risk of thromboembolic complications or adverse clinical outcomes. Long-term follow-up shows outcomes similar to controls with dual-chamber pacemakers.
Area of Science:
- Cardiology
- Medical Devices
- Thrombosis
Background:
- Transvenous leads are crucial for cardiac device function.
- Multiple leads may increase the risk of thromboembolic complications.
- The clinical significance of multiple noninfected transvenous leads requires further investigation.
Purpose of the Study:
- To prospectively evaluate the incidence and clinical significance of thromboembolic complications in patients with multiple (≥3) noninfected transvenous leads.
- To compare clinical outcomes with age-matched controls having dual-chamber (DDD) pacemakers.
Main Methods:
- Prospective evaluation of 48 consecutive patients with ≥3 noninfected transvenous leads.
- Clinical follow-up included assessment of venous thrombosis, right heart failure, hospital admissions, and mortality.
- Annual echo-Doppler studies evaluated cardiac structures and function.
- Comparison with age-matched controls with DDD pacemakers.
Main Results:
- No significant differences in clinical thromboembolic complications, hospital admissions, or mortality were observed compared to controls.
- Transient venous thrombosis occurred in 17% of the study group versus 15% in controls.
- Increased tricuspid regurgitation was noted in patients with multiple ventricular leads (24% vs 4% in controls), but without clinical signs of right heart failure.
Conclusions:
- Patients with multiple (≥3) noninfected transvenous leads do not experience adverse clinical outcomes during long-term follow-up.
- The presence of multiple leads does not increase the risk of thromboembolic complications or mortality compared to standard dual-chamber pacemaker patients.
- While echocardiographic changes like tricuspid regurgitation may be more frequent, they do not translate to clinical right heart failure.
Abstract:
To prospectively assess the incidence and clinical significance of thromboembolic complications in patients with multiple (> or = 3) noninfected transvenous leads; 48 consecutive patients were evaluated. Half of the patients had two ventricular leads and one atrial lead, 15 patients had two atrial leads and one ventricular lead, while 9 patients had two ventricular and two atrial leads. No additional care was provided except for aspirin (80 mg bid) and annually performed echo-Doppler studies. Clinical follow-up included signs and symptoms of subclavian and/or axillary vein thrombosis, the presence of right congestive heart failure, the number of hospital admissions, and death. Echo-Doppler studies assessed the presence of an enlarged right atrium or ventricle, right atrial or ventricular spontaneous contrast, and the presence of tricuspid regurgitation. During a total follow-up of 7.4 +/- 2.2 years there were no differences in the incidence of clinical variables as compared to age-matched controls with DDD pacemakers. The most common complication was transient venous thrombosis (mostly presenting as venous prominence 1-2 weeks after implantation), which was seen in 17% of the study group versus 15% in controls (NS). Cumulative mortality was not different in both groups (13% in the study group vs 15% in controls). No differences were present with respect to hospital admissions (1.1 +/- 0.27/year in the study group vs 1.2 +/- 0.30/year in the controls). In patients with multiple ventricular leads, tricuspid regurgitation on echo-Doppler studies was more frequent (24%) as compared to controls (4%); however, clinical signs of right heart failure were equally distributed. Thus, patients with multiple (> or = 3) noninfected leads have no clinical adverse outcome during long-term follow-up.
Related Concept Videos
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Cardiac Catheterization III: Left Heart Catheterization
Acute Coronary Syndrome III: Diagnostic Studies