Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
A quality control method for detecting energy changes of medical accelerators
1Department of Radiation Oncology, Emory University School of Medicine, Atlanta, GA 30322, USA. patton@radonc.emory.org
Abstract:
A description is presented of a simple and sensitive method for detecting a change in the energy of the electrons bombarding the target of medical accelerators. This technique is useful for x-ray beams with end point energy in the range of 15.7 to 25 MeV. The method is based on the photoactivation of 160 and 14N in a small sample of ammonium nitrate. It was found that the ratio of the activity induced in the oxygen divided by that produced in the nitrogen can be used as a quality control technique to detect a change in the energy of the electrons that bombard the target of the accelerator. An electron energy change of the order of 0.2 MeV can be determined using this method.