Related Experiment Videos
[Electrode problems in pacemaker wearers (experiences with three electrode types) (author's transl)]
Abstract:
3 types of electrodes (Medtronic 6903, EMT 588, MIP 2000) of 123 patients are examined by following criterial 1. rate of dislocation, 2. development of threshold, 3. time of application. We found: The Medtronic has the smallest rate of dislocation, the lowest range of threshold and is quickly placed in the right ventricle. The rate of dislocation of the EMT 588 is still just acceptable and it also has a sufficient low threshold, but in comparison to the other helical-wire electrodes the disadvantage of being rather difficult to place in the right ventricle. The MIP 2000 has not proved to be useful due to its large rate of dislocation.