Hemodynamic changes and right heart support during vertical displacement of the beating heart
1Department of Cardiothoracic Surgery, Carmel Medical Center, Technion IIT, Haifa, Israel. docporat@internet-zahav.net
Background:
Hemodynamic instability during heart displacement in off-pump multivessel coronary artery bypass grafting might be related to right heart dysfunction. The Enabler (HemoDynamics Systems Ltd, Upper Yoqneam, Israel) is a cannula pump that expels blood from the right atrium into the pulmonary artery. We studied the hemodynamic changes and the role of the enabler during heart displacement.
Methods:
Nine anesthetized sheep were assessed for hemodynamic changes during 90-degree heart displacement with or without Enabler support. Hemodynamic parameters included cardiac output, systemic arterial blood pressures, and left and right heart filling pressures.
Results:
Heart displacement caused a significant decrease in cardiac output and systemic blood pressure (46%+/-5%, p = 0.001; and 20%+/-5%, p = 0.009, respectively), with a concomitant 137%+/-24% (p = 0.003) increase in central venous pressure. No significant change in left atrial pressure was observed. Activation of the Enabler caused a significant increase in cardiac output and systemic blood pressure (67%+/-15%, p = 0.01; and 17%+/-7%, p = 0.04, respectively), as well as a decrease in central venous pressure by 49%+/-8% (p = 0.0001).
Conclusions:
Heart displacement causes hemodynamic instability mainly by right heart dysfunction. The Enabler significantly stabilized circulation during vertical displacement of the beating heart.