Aeration time following ethylene oxide sterilization for reusable rigid sterilization containers: concentration of
S Nakata1, K Umeshita, H Ueyama
1Surgical Center, Osaka University Medical School, Japan. nakata@hp-op.med.osaka-u.ac.jp
Abstract:
Because ethylene oxide (EO) gas is toxic to humans, restrictions have been imposed on its use for sterilization, specifying allowable levels of residual EO remaining in sterilized apparatus and materials. However, the aeration time that optimizes the removal of the remaining EO when a rigid sterilizing container is used for a vessel had not been identified. Therefore, polyvinyl chloride, which easily adsorbs EO, was placed in rigid sterilizing containers, and aeration was carried out after 1, 8, 12, 17, and 24 hours. After standard EO sterilization, the EO concentrations remaining in the air in the rigid containers were measured. The results indicate that a period of 17 hours of aeration is appropriate when a rigid sterilizing container is used.
Related Concept Videos
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Bioreactor Controls-II
Autoxidation of Ethers to Peroxides and Hydroperoxides
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Carboxylic Acids to Methylesters: Alkylation using Diazomethane