Related Experiment Video
Updated: Apr 23, 2026

Experimental and Imaging Techniques for Examining Fibrin Clot Structures in Normal and Diseased States
Published on: April 1, 2015
Thrombus on indwelling central venous catheters: the histopathology of "Fibrin sheaths"
J N Suojanen1, D P Brophy, I Nasser
1Department of Radiology, Beth Israel Deaconess Medical Center, 1 Deaconess Road, Boston, MA 02215, USA.
Insights
Central venous catheters (CVC) used for dialysis often fail due to thrombus formation. This blockage can occur with or without a protein coating, commonly called a fibrin sheath.
Area of Science:
- Nephrology
- Vascular Access
- Biomedical Engineering
Background:
- Central venous catheters (CVCs) are crucial for dialysis.
- Catheter failure is a significant clinical challenge.
- Fibrin sheaths are often cited as the primary cause of CVC dysfunction.
Purpose of the Study:
- To investigate the histopathology of material occluding dialysis catheter ports.
- To determine the composition of "fibrin sheaths" removed from dialysis catheters.
Main Methods:
- Ten dialysis catheter strippings were performed.
- Gross and microscopic examination of the removed catheter material.
Main Results:
- Histologic analysis revealed the presence of thrombus.
- Thrombus was observed both with and without an associated proteinaceous sheath.
- The findings challenge the sole attribution of failure to fibrin sheaths.
Conclusions:
- Dialysis catheter failure is primarily attributed to thrombus formation.
- Thrombus can develop irrespective of a protein coating (fibrin sheath).
- Understanding the role of thrombus is key to preventing dialysis catheter dysfunction.
Purpose:
Central venous catheters (CVC) may fail for many reasons, though "fibrin sheaths" blocking catheter ports are usually implicated. We examined the sheaths removed from dialysis catheters to determine their histopathology.
Methods:
Ten catheter strippings were performed and the removed material was studied grossly and microscopically.
Results:
The histologic specimens showed thrombus both with and without a proteinaceous sheath.
Conclusion:
Dialysis catheters fail because of thrombus formation. This can occur in either the absence or presence of a protein coating on the catheter, the so-called "fibrin sheath."
Related Concept Videos
Anticoagulant Drugs: Low-Molecular-Weight Heparins
Extrinsic and Intrinsic Pathways of Hemostasis
The Extrinsic Pathway
The extrinsic pathway of coagulation is typically initiated by tissue damage that exposes blood to tissue factor (TF), a protein released by the damaged tissue cells outside the blood vessels—this interaction with TF triggers biochemical reactions involving specific clotting factors. The key player here is Factor VII, which...
Clot Retraction and Fibrinolysis
Venous Thrombosis I: Introduction
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies
Venous Thrombosis III: Interprofessional Care