Efficacy of a washer-pasteurizer for disinfection of respiratory-care equipment
W A Rutala1, D J Weber, M F Gergen
1Department of Medicine, University of North Carolina, School of Medicine, Chapel Hill 27599-7030, USA.
Abstract:
We evaluated the efficacy of a commercial washer-pasteurizer. Carriers were inoculated with 10(4) to 10(6) test organisms and pasteurized at 170 degrees F for 30 minutes. Pasteurization eliminated all test organisms with the exception of Bacillus subtilis spores. Pasteurization appears efficacious for the disinfection of respiratory-care equipment and could result in a cost savings of approximately $30,000 per year.
Related Concept Videos
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Physical Methods for Controlling Microbial Growth: Radiation and Filtration
Physical Methods for Controlling Microbial Growth: Temperature
Pasteurization and Food Preservation