False alarms and sensitivity of conventional pulse oximetry versus the Masimo SET technology in the pediatric
S Malviya1, P I Reynolds, T Voepel-Lewis
1Department of Anesthesiology, The University of Michigan Health System, Ann Arbor, Michigan 48109-0211, USA. smalviya@umich.edu
Insights
Masimo Signal Extraction Technology (Masimo SET) reduces false alarms and identifies more true alarms in pediatric postanesthesia care compared to conventional pulse oximetry (CPO). This technology improves SpO2 monitoring accuracy in children.
Area of Science:
- Biomedical Engineering
- Anesthesiology
- Pediatric Critical Care
Background:
- Pulse oximetry is crucial for monitoring oxygen saturation (SpO2) in postanesthesia care units (PACU).
- Conventional pulse oximetry (CPO) can generate false alarms (FA) and miss true alarms (TA), impacting patient care.
- Masimo Signal Extraction Technology (Masimo SET) is designed to improve SpO2 measurement accuracy, especially during patient motion.
Purpose of the Study:
- To compare the incidence and duration of false alarms (FA) and the sensitivity for true alarms (TA) between CPO and Masimo SET in pediatric PACU patients.
- To evaluate the impact of Masimo SET on data dropout rates.
Main Methods:
- A prospective study involving 75 children in the PACU.
- Synchronous SpO2 data acquisition using both CPO and Masimo SET sensors placed on separate digits.
- Documentation of patient motion, respiratory events, and clinical observations to define false alarms (data dropout, false negative, artifactual SpO2) and true alarms.
Main Results:
- Masimo SET identified significantly more true alarms (100%) compared to CPO (59%) (P < 0.05).
- CPO generated twice the number of false alarms (10 vs 4) compared to Masimo SET (P < 0.05).
- Incidence and duration of data dropouts were similar between the two technologies.
Conclusions:
- Masimo SET demonstrates an advantage over CPO in pediatric PACU by reducing false alarms and increasing true alarm detection.
- This improved accuracy in SpO2 monitoring may lead to better clinical decision-making and patient outcomes.
Unlabelled:
We compared the incidence and duration of false alarms (FA)and the sensitivity of conventional pulse oximetry (CPO) with Masimo Signal Extraction Technology (Masimo SET; Masimo Corporation, Irvine, CA) in children in the postanesthesia care unit. Disposable oximeter sensors were placed on separate digits of one extremity. Computerized acquisition of synchronous data included electrocardiograph heart rate, SpO(2), and pulse rate via CPO and Masimo SET. Patient motion, respiratory, and other events were simultaneously documented. SpO(2) tracings conflicting with clinical observations and/or documented events were considered false. These were defined as 1) Data dropout, complete interruption in SpO(2) data; 2) False negative, failure to detect SpO(2) = 90% detected by another device or based on observation/intervention; 3) FA, SpO(2) = 90% considered artifactual; and 4) True alarm (TA), SpO(2) = 90% considered valid. Seventy-five children were monitored for 35 +/- 22 min/patient (42 h total). There were 27 TAs, all of which were identified by Masimo SET and only 16 (59%) were identified by CPO (P < 0.05). There was twice the number of FAs with CPO (10 vs 4 Masimo SET; P < 0.05). The incidence and duration of data dropouts were similar between Masimo SET and CPO. Masimo SET reduced the incidence and duration of FAs and identified a more frequent incidence of TAs compared with CPO.
Implications:
Pulse oximetry that incorporates Masimo Signal Extraction Technology (Masimo Corporation, Irvine, CA) may offer an advantage over conventional pulse oximetry by reducing the incidence of false alarms while identifying a higher number of true alarms in children in the postanesthesia care unit.
More Related Videos
07:27How to Administer Near-Infrared Spectroscopy in Critically ill Neonates, Infants, and Children
Published on: August 19, 2020
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...